Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 30: (1 year) CAL VER /5700,5725 DATE: 31-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 92 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke Model 30 Clamp Meter Service Information (PN 929042 January 1994). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 57_30.bmp # 57_30_CW.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP This procedure requires a 50-turn Toroid Coil. 1.004 DISP Refer to the Model 30 Clamp Meter Service Information 1.004 DISP for directions for constructing a suitable coil. 1.005 HEAD {CURRENT} 1.006 PIC 57_30 1.007 HEAD CURRENT: {200A Range} 1.008 RNG 200A AC 1.009 5700 0.4A 60H S 2W 1.010 ACC 20.0A 650P% 0.002U 1.011 MEMI Enter UUT reading in amps. 1.012 MEME 1.013 MEMC 200 A 1.2U 60H 2.001 HEAD CURRENT: {400A Range} 2.002 RNG 400A AC 2.003 5700 0.8A 60H S 2W 2.004 ACC 40A 650P% 0.002U 2.005 MEMI Enter UUT reading in amps. 2.006 MEME 2.007 MEMC 400 A 4U 60H 3.001 HEAD CURRENT: {200A Range} 3.002 PIC 57_30_CW 3.003 RNG 200A AC 3.004 5700 3.8A 60H B1 S 2W 3.005 ACC 190.0A 400P% 0.0085U 3.006 MEMI Enter UUT reading in amps. 3.007 MEME 3.008 MEMC 200 A 2.7U 60H 4.001 HEAD CURRENT: {400A Range} 4.002 RNG 400A AC 4.003 5700 7A 60H B1 S 2W 4.004 ACC 350A 400P% 0.0085U 4.005 MEMI Enter UUT reading in amps. 4.006 MEME 4.007 MEMC 400 A 8U 50H 5.001 5700 8A 60H B1 S 2W 5.002 ACC 400A 400P% 0.0085U 5.003 MEMI Enter UUT reading in amps. 5.004 MEME 5.005 MEMC 400 A 12U 50H 6.001 HEAD VOLTAGE: {200V Range} 6.002 5700 200 20.0V 0.5U 60H 2W 7.001 5700 200 190.0V 2.6U 60H 2W 8.001 HEAD VOLTAGE: {600V Range} 8.002 5700 600 60V 4U 60H 2W 9.001 5700 600 600V 10U 60H 2W 10.001 HEAD {RESISTANCE} 10.002 5700 200 190Z 3.1U CW 11.001 END